PROCEDURE :
|
|||||||||||||||||||
PRECAUTIONS:
Place
the instrument as far as possible away from any strong magnetic or electric
field or any electrical apparatus generating a high frequency.
Avoid
placing it where severe vibrations exist or when the sun beats.
Place
the instrument free from dust and corrosive gases.
To
extend the life of the source lamp, have them powered only when required.
Cells
should be cleaned and handled with great care. Wipe the cell properly with
tissue or dry soft cloth.
Never
use alkali, abrasive / etching material or hot concentrated acids.
|
|||||||||||||||||||
OPERATIONS:
Switch
on main switch.
When
the instrument power is turned ON, UV-Spectrophotometer the starts
executing various checks and initializations.
|
|||||||||||||||||||
CALIBRATION:
CONTROL
OF WAVELENGTHS:
Preparation
Of 1.4 M Perchloric Acid Solution:
Take
11.5 ml of Perchloric acid (about 60%w/w strength) into a clean
100 ml volumetric flask. Dilute to the volume with distilled water(Note
:Dilute 8.5 ml to 100 ml with distilled water in case of 70% Perchloric
acid)
Preparation
Of Holmium Perchlorate (With Holmium Oxide):
Weigh
accurately about 400 mg Holmium oxide.
Transfer into a 10 ml volumetric flask containing 5 ml of 1.4 M Perchloric acid solution. Shake the solution well to dissolve Holmium oxide and make up the volume to the mark with 1.4 M Perchloric acid solution.
Procedure
For Control Of Wavelengths:
Take a
pair of cuvette having a path length of 1 cm.
Perform
the base line correction with 1.4 M Perchloric acid solution from 200 to 600
nm.
Remove the cell from sample compartment. Rinse the sample cell with Holmium perchlorate solution, fill the cell with the same, clean the smooth surfaces of the cell with tissue paper and place it in the sample compartment.
|
|||||||||||||||||||
CONTROL OF ABSORBANCE:
Preparation
Of 0.005 M Sulphuric Acid:
a)
Take a clean 1000 volumetric flask and fill to the mark with distilled water
by adding 0.27 ml Conc. Sulphuric acid to the above flask.
b)
b) Take a clean 100 volumetric flask and fill to the mark with distilled water by adding 0.027 ml Conc. Sulphuric acid to the above flask.
Preparation
Of K2Cr2O7 Solution:
Take Analytical Reagent grade K2Cr2O7 (potassium dichromate) in to a
weighing
bottle and heat at130oC in a laboratory oven for 2 hours and
Cool it in a desiccator.
Solution A: weigh accurately quantity not less than 57.0 mg and not more than 63.0 mg and dissolve in 1000 ml volumetric flask containing 300 ml of 0.005 M Sulphuric acid. Shake the flask well to dissolve Potassium dichromate completely. Make up the volume with 0.005 M Sulphuric acid.
Solution
B : weigh accurately quantity not less than 57.0 mg and not
more than 63.0 mg and dissolve in 100 ml volumetric flask
containing 50 ml of 0.005 M Sulphuric acid. Shake the flask well
to dissolve Potassium dichromate completely. Make up the volume with 0.005 M
Sulphuric acid.
Procedure
For Control Of Absorbance:
Take a pair of cuvette having a path length of 1 cm. Perform base line correction with 0.005 M Sulphuric acid from 200 to 600 nm. Remove the cell from sample compartment. Rinse the sample cell with solution A, fill the cell with the same, clean the smooth surfaces with tissue paper. Place the cell in sample compartment, scan from 200 to 600 nm and measure the absorbances at 235 nm, 257 nm, 313 nm and 350 nm. Rinse the sample cell with solution B, fill the cell with the same, clean the smooth surfaces with tissue paper. Place the cell in sample compartment, measure the absorbances at 430 nm.
Acceptance
Criteria:
Calculate
the specific absorbance using the following equation.
Specific absorbance at x nm = Absorbance at x nm X 1000 X 10 weight of sample(mg) The specific absorbances against each wavelength should be as follows :
|
|||||||||||||||||||
LIMIT OF STRAY LIGHT:
Preparation
Of 1.2% W/V Potassium Chloride Solution
Take
about 3 gm of Potassium chloride dry at 250oC for about 1 hours.
Weigh accurately about 1.2 g of Potassium chloride. Transfer into a 100 ml volumetric flask containing 75 ml distilled water. Shake the flask well to dissolve completely and make up to the volume with water.
Procedure for limit of stray light
Take a
pair of cuvette having a path length of 1 cm. perform the auto zero with
distilled water.
Remove
the cell from sample compartment. Rinse the cell in the potassium
chloride solution, fill the cell with the same, clean the smooth
surfaces with tissue paper and place it in sample compartment.
Measure the absorbance at 200 nm.
Acceptance Criteria
The
absorbance should be more than 2.
|
|||||||||||||||||||
RESOLUTION POWER:
Preparation
of 0.020 % v/v of toluene in Hexane solution:
Dilute
2 ml of Toluene to 200 ml with Hexane. Further dilute 2 ml of above
solution to 100 ml with Hexane.
Record
the spectrum in the range 260 to 275 nm of 0.020 % v/v toluene in Hexane
using Hexane in the reference cell. Calculate the ratio of the absorbance at
the maximum at about 269 nm to that at the minimum at about 266 nm
Acceptance
Criteria : Ratio
is not less than 1.5.
|
|||||||||||||||||||
SPECIFICATION
OF CELLS:
In
`Photometry mode ’ Select transmittance mode. Set the wavelength at 200 nm.
Then press `auto zero’ to get 100 % transmittance with air blank in both the
cell holders.
Then
place the cell in the sample compartment filled with distilled water using
air blank and read the transmittance. Similarly repeat the operation
for 220 nm and 240 nm.
Repeat
the test for second cell.
Acceptance
Criteria: The
transmittance values should meet the requirement as given in the following
table.
|
|||||||||||||||||||
LINE FLATNESS:
With
air blank in the sample and reference compartment, perform the
test.
Scan
in the wavelength range 220 to 700 nm.
Obtain
the spectrum in the % transmittance mode.
Acceptance
Criteria:
The
line obtained should not deviate from a horizontal line at any point by more
than 98.0% - 102.0% transmittance.
|
|||||||||||||||||||
SOLVENTS:
Take
the absorbance of following solvents at about 254 nm with reference to
distilled water.
a) Ethanol (95%)
b) Ethanol
c) Methanol
d) Cyclohexane
Acceptance
Criteria: Absorbance
should not exceed 0.10
|
Wednesday 4 October 2017
Calibration of UV-Spectrophotometer (2017)
Posted by mypharmasolution.blogspot.com on October 04, 2017