Calibration of Friability Test Apparatus





















PROCEDURE :

PRECAUTION
Handle the acrylic drum with care.

Do not try to remove the drum when instrument is ON.

Tighten the knob of the drum properly before starting the Friability Apparatus.

Align the drum marking with that of screw to avoid the error in counting.



OPERATION:
Weigh tablets equivalent to 6.5 gm whose average weight is equal to or less than 650 mg. and note the weight in calculation sheet.

Take 10 tablets whose average weight is more than 650 mg. and note the weight in calculation sheet.

Remove the acrylic cover of the drum and place the tablets inside the drum.

Replace the cover and tighten with the help of plastic knob.

Switch the MAINS power ON.

Ensure the digital display indicates 0000, if not - press Reset.

To set the required rotation, Press ’SET’ switch.

Yellow LED will glow.

Set the rotation by ‘UP’ & ‘DOWN’ switch provided on the front side of the Friability Apparatus.

Press START switch after rotation is set.

         a)  Green coloured LED will glow and yellow LED will turn OFF.

         b)  Machine starts with rotation of drum.

         c)  Digital display increments with each rotation till equal the set number.    
         d)  Machine stops when set and displayed rotation become equal.

         c)  Alarm will sounds and red coloured LED will glow.

         e)  Reset alarm with STOP switch.

Remove the tablet from the drum, weigh and note the weight in calculation sheet.

Remove the broken tablets and do not consider for weighing.

Calculate the friability (loss of weight in percent).

Friability  =                 (A - B) x 100           
                                             A              
    Where ,
                A = initial weight of tablets in gm.
                B = weight of unbroken tablets in gm after the  rotation completed.                            

Acceptance criteria for friability of uncoated tablet are not more than 1.0%.

If obviously cracked, chipped or broken tablets are present in the sample after rotation, the sample fails the test.

Inform to Quality Control Head in case of sample fails the test.



CALIBRATION PROCEDURE :

Ensure that tablet friability apparatus is clean and free from dust.

Switch “ON” the main supply.

Set the 100  rotation as mentioned above.

Press START switch.

Observed the rotation per mentioned minutes using calibrated stop watch.

Note the observed rotation in format for calibration of  Friability Apparatus.



ACCEPTANCE CRITERIA

The rotations of drums observed manually should be within tolerance limit.



Sr. No.

Time in minutes

Rotations set

Tolerance

1
1.0
25
+ 1

2
2.0
50
+  2

3
4.0
100
+ 4